1
/
of
1
日置電機
Digital multimeter DT4252 x 1
Digital multimeter DT4252 x 1
Regular price
¥24,200
Regular price
Sale price
¥24,200
Unit price
/
per
Shipping calculated at checkout.
Couldn't load pickup availability
Speed up your professional work! High safety and high-speed response standard series
DCV basic accuracy of ±0.3%, ACV frequency characteristic of 40 to 1kHz, standard type
Low-pass filter 100Hz/500Hz cuts harmonics
Equipped with multiple functions such as C measurement and frequency measurement
USB communication function for PC measurement (optional)
Wide operating temperature range: -10℃ to 50℃ Dimensions: 84W x 174H x 52D mm
Weight: 390g (including battery and holster)
Accuracy guaranteed for 1 year, post-adjustment accuracy guaranteed for 1 year
DC voltage
Range: 600.0mV to 1000V, 5 ranges
Basic accuracy: ±0.3%rdg.±5dgt.
AC voltage
Range: 40-500Hz 6.000V-1000V, 4 ranges (frequency response: 40Hz-1kHz)
Basic accuracy: ±0.9% rdg. ±3 dgt. (RMS rectification, crest factor 3 or less)
DC current
Range: 6.000A/10.00A, 2 ranges
Basic accuracy: ±0.9%rdg.±5dgt.
Alternating current
Range: 40-500Hz 6.000A/10.00A, 2 ranges
Basic accuracy: ±1.4% rdg. ±3 dgt. (RMS rectification, crest factor 3 or less)
resistance
Range: 600.0Ω to 60.00MΩ, 6 ranges
Basic accuracy: ±0.7%rdg.±5dgt.
Capacitance
Range: 1.000μF to 10.00mF, 5 ranges
Basic accuracy: ±1.9%rdg.±5dgt.
frequency
Range: ACV/ACA: 3 ranges from 99.99Hz (5Hz or higher) to 9.999kHz, ACV only: 99.99kHz
Basic accuracy: ±0.1%rdg.±1dgt.
Continuity check
Continuity ON threshold: 25Ω or less (buzzer sounds, red LED lights up)
Continuity OFF threshold: 245 Ω or more
Response time: 0.5 ms or faster
Diode Test
Open circuit voltage: 5.0V or less
Measurement current: 0.5mA
Forward threshold: 0.15V to 1.5V
Other functions: Filter, display value hold, auto hold, maximum/minimum/average value display, relative display, auto power save, USB communication (optional)
Display: 4-digit LCD display, max. 6000 dgt., bar graph
Display update rate: 5 times/s (Capacitance: 0.05 to 5 times/s depending on the measured value, Frequency: 1 to 2 times/s, Temperature: 1 time/s)
Power supply: 4 AAA alkaline batteries (LR03)
Continuous use time: 130 hours (when backlight is off)
Accessories: Test lead L9207-10 x 1, Holster x 1, Instruction manual x 1, AAA alkaline batteries (LR03) x 4
*About DMM accuracy
Since DMM has many measurement functions and ranges, only typical values are listed as basic accuracy. Please refer to the PDF catalog for accuracy for each range.
View full details