It boasts precise temperature control and the durability of the stainless steel coil.
The cooling capacity has been improved by adopting a high-performance refrigerator.
The circulating water cooler uses a square coil made of SUS304, so it is compatible with pure water. (*If using pure water, please use a liquid with an electrical conductivity of 10μs/cm or more (low purity).)
The refrigerant used is R-404, which has an ozone depletion factor of 0, making it environmentally friendly.
The external circulation nozzle is installed on the front of the controller and can be freely rotated 180°.
The temperature setting display uses a two-stage PV/SV display, allowing you to check the actual value and set value at a glance.
By attaching the SB (support rod and bracket set), flasks and external sensors placed in the tank can be fixed with clamps.
Type: -30 to +80
Cooling capacity(W): 340
Temperature setting and display method: Key input method (0.1℃), PV/SV digital two-level display
Temperature control method: Microcomputer PID control
Temperature sensor: Resistance thermometer (Pt100Ω)
Temperature accuracy: ±0.1℃ (display accuracy: may exceed range depending on conditions)
Heater: 1kW
External dimensions: 305 x 490 x 700 mm
Tank interior dimensions: 239 x 299 x 200 mm
Effective opening dimensions: 235 x 162 x 200 mm
Tank capacity: Approx. 12L
Material: Water tank and cooling coil: Stainless steel (SUS304)
Types of liquid used: Pure water, clean water, ethylene glycol, ethyl alcohol
Refrigerant: R-404A
Circulation pump (50/60Hz): Maximum flow rate 6.7/7.5L/min, Maximum head 3.5/5.0m
Circulation nozzle outer diameter: φ11mm (suitable tube inner diameter φ8mm)
Safety devices: Self-diagnosis function (detection of upper and lower temperature limits, float type dry-burn prevention, sensor abnormality, heater disconnection, leakage current/overcurrent breaker, refrigerator overload relay)
External input/output: External temperature sensor input terminal (for Pt100), analog output meter for thermometer (1 to 5V), alarm output terminal (DC12 to 17V, 20mA)
Timer function: 1 minute to 48 hours (in 1 minute increments)
Power supply: AC100V 50/60Hz
Maximum operating current: 15A
Weight: 39kg
Accessories: 2 external circulation nozzles, 1 outlet adapter, 2 insulated hoses (inner diameter φ8 mm x 1 m)
*The freezer will not work if the liquid temperature is over 40°C.
*Please avoid continuous operation at high temperatures (around 80°C).
*Circulation pump performance is measured when the pump is used alone.
This product can be provided with validation services (IQ/OQ). Please feel free to contact us.
*Validation service refers to the performance of installation qualification (IQ) and operational qualification (OQ) at the customer's site and the submission of an inspection report.
View full details